Purpose
The door-ajar warning light notifies the driver that any door or the liftgate is open.
Function
When the instrument cluster receives any of the following signals sent from the rear body control module (RBCM) via the CAN signal, the door-ajar warning light illuminates.
Door latch switch ON signal (door open)
Liftgate latch switch ON signal (liftgate open)
Construction
The instrument cluster controls illumination on/off of the door-ajar warning light using the instrument cluster microcomputer based on the door latch or liftgate latch position signal from the rear body control module (RBCM).
The door-ajar warning light is set in the instrument cluster.

Operation
1. The instrument cluster receives (1) the door latch switch ON signal or liftgate latch switch ON signal from the rear body control module (RBCM).
2. The instrument cluster microcomputer turns the transistor on (2) based on each signal.
3. When the transistor turns on, a ground circuit with the door-ajar warning light is established and the door-ajar warning light illuminates (3).

Fail-safe
Function not equipped.
Cruise Set Indicator Light (Green)
Engine Oil Pressure Warning LightClock Spring Adjustment [Standard Deployment Control System]
1. Set the front wheels straight ahead.
CAUTION:
The clock spring will break if over?wound. Do not forcibly turn the clock
spring.
2. Turn the clock spring clockwise until it stops.
3. Turn the clock spring counterclockwise approx. 2 turns.
4. Align the mark on the clo ...
Brake Pedal
Purpose/Function
An intrusion-minimizing brake pedal has been adopted to the brake pedal to
provide for a measure of safety in the event of an accident.
The intrusion-minimizing brake pedal mechanism reduces impact to the lower
extremities of the driver by minimizing the amount ...
Pressure Control Valve Inspection [Skyactiv G 2.0]
1. Remove the pressure control valve assembly..
CAUTION:
Do not apply a pressure 36 kPa {0.37 kgf/cm2, 5.2 psi} or more. Doing so
will damage the pressure control valve and evaporative hose.
2. Apply a pressure of 36 kPa {0.37 kgf/cm2, 5.2 psi} to port A.
3. Verify that ther ...