U.S.A. And CANADA
Leakage inspection
1. Perform the following procedures:
Do not operate CV solenoid valve.
Assemble the hose used for inspection to port A.
Cover ports B and C with the caps.

CAUTION:
Do not apply a pressure 20 kPa {150 mmHg, 5.91 inHg} or more to the charcoal canister for 1 min. or more. Doing so will damage the charcoal canister.
2. Apply pressure (19.6 kPa {147 mmHg, 5.79 inHg}
) through the hose, and verify that air does not leak from the charcoal canister.

If there is airflow, replace the charcoal canister..
If there is no airflow, perform the following “Airflow Inspection”..
Airflow inspection
1. Perform the following procedure to verify the charcoal canister airflow.
Plug the port B.

Apply a pressure of 1.8 kPa {14 mmHg, 0.53 inHg}
to port A.
Verify that there is airflow from port C.
If there is no airflow, replace the charcoal canister..
Plug the port A.

Apply a pressure of 2.0 kPa {15 mmHg, 0.59 inHg}
to port C.
Verify that there is airflow from port B.
If there is no airflow, replace the charcoal canister..
Except U.S.A. And CANADA
Leakage inspection
1. Perform the following procedures:
Assemble the hose used for inspection to port A.
Cover ports B and C with the caps.

CAUTION:
Do not apply a pressure 20 kPa {150 mmHg, 5.91 inHg} or more to the charcoal canister for 1 min. or more. Doing so will damage the charcoal canister.
2. Apply pressure (19.6 kPa {147 mmHg, 5.79 inHg}
) through the hose, and verify that air does not leak from the charcoal canister.

If there is airflow, replace the charcoal canister..
If there is no airflow, perform the following “Airflow Inspection”..
Airflow inspection
1. Perform the following procedure to verify the charcoal canister airflow.
Plug the port B.

Apply a pressure of 0.98 kPa {7.4 mmHg, 0.29 inHg}
to port A.
Verify that there is airflow from port C.
If there is no airflow, replace the charcoal canister..
Plug the port A.

Apply a pressure of 0.98 kPa {7.4 mmHg, 0.29 inHg}
to port C.
Verify that there is airflow from port B.
If there is no airflow, replace the charcoal canister..
Catch Tank
Charcoal Canister Removal/InstallationLiftgate Latch And Lock Actuator Inspection
The following actuator and switch are integrated with the liftgate latch
and lock actuator.
From the actuator/switch name and operation in the following table, select
the actuator or switch to be inspected, and perform the inspection following
the inspection procedure desc ...
Energy Absorbing Form Removal/Installation
1. Disconnect the negative battery cable..
2. Remove the seal board upper..
3. Remove the front bumper..
4. Insert the flathead screwdriver shown in the figure, push fasteners A for
back side of front bumper reinforcement arrow shown in the figure.
5. Pull the energy-absorbing form in ...
Starting the Engine
WARNING
Radio waves from the key may affect medical devices such as pacemakers: Before
using the key near people who use medical devices, ask the medical device manufacturer
or your physician if radio waves from the key will affect the device.
NOTE
• The key must be carried because the
ke ...